Measuring the Oxygen Content of Blood with LEDs

Pulse oximetry is a non-invasive method for determining haemoglobin (Hb) saturation through oxygen in the blood and heart frequency. It uses the fact that oxygen-rich and oxygen-poor haemoglobin absorb red and infrared light differently. In order to measure the oxygen content reliably, LEDs with…

Published by LED professional ·

LED professional

Pulse oximeters can use two different measurement methods to detect the oxygen content in the blood: transmissive oximetry passes two wavelengths of light through a body part, for example a fingertip or earlobe. Reflective oximetry is used in fitness trackers or smart watches. Both methods measure the oxygen content in the blood on the basis of two basic principles: First, they distinguish between oxygen-rich oxyhaemoglobin (HbO2) and oxygen-poor deoxyhaemoglobin (Hb), and then they detect the oxygen saturation only in the arterial blood.

Different absorption spectra of the blood components

Pulse oximeters take advantage of the fact that red and infrared light is absorbed differently by oxygen-rich and oxygen-poor blood: HbO2 absorbs more infrared light and less red light than Hb. Due to the higher reflection of red light it appears bright red. By contrast, Hb absorbs more red light and thus appears darker.

Transmissive pulse oximetry usually uses a red LED in the range of about 650 nm and an IR LED in the range of about 900 nm. Both LEDs screen a fingertip and a phototransistor on the other side captures the relative amount of absorbed red and IR light. A micro-controller then determines the proportion of Hb and HbO2 and calculates the oxygen content from the measured parameters. In fact, only the arterial blood is of interest. With pulse oximetry it is possible to detect the oxygen content only of arterial blood. This is based on the fact that the amount of absorbed red and infrared light varies with every heartbeat due to the pulsating changes in blood volume. By contrast, veins, bones and tissue show relative constant absorption characteristics.

Reliable measurements need the right components

The right components are essential to reliably determine the oxygen content of blood. The LEDs and photodetectors that are used are particularly important. Würth Elektronik's red LED 150224SS73100 is characterized by an exact wavelength of 652 nm and high efficiency. With a current of IF=20mA a typical luminous intensity of 130 mcd is achieved. In combination with a good phototransistor this results in a high signal-to-noise ratio. In order to obtain a good signal-to-noise ratio with the IR LED as well, it should have about twice the luminous intensity of the red LED in this application. The LED 15414194BA210 from the WL-SITW-SMT series is particularly suitable here, as it has a very high efficiency and a high radiation intensity. The phototransistor WL-STTW SMT, 1541411NBA2010 is recommended as light sensor. Although the component has the highest light sensitivity at a wavelength of 940 nm, it also detects light signals with a wavelength of 660 nm. Therefore, only one single phototransistor is needed for the measurement. The perfect emitter-detector combination from Würth Elektronik for pulse oximetry and other IR applications is available from stock in any quantity.

Further information can be found at we-online.de/web/en/wuerth_elektronik/start.php, katalog.we-online.de/en/led/WL-SITW and katalog.we-online.de/en/led/WL-STTW

Did you like this article?

Share this article

You could also like this

UV / IR

Far-UVC LEDs transmit data at record speed | Ferdinand-Braun-Institut

Researchers at Ferdinand-Braun-Institut (FBH), the University of Strathclyde, and the University of Cambridge demonstrated wireless optical communication using far-UVC LEDs at wavelengths below 235 nm, achieving data rates up to 1.5 gigabits per second over a 30 cm line-of-sight link. The FBH-developed LEDs are optimized for high optical power and fast modulation by segmenting the emitting surface to reduce capacitance and increase current density. Far-UVC offers low solar background outdoors and stronger Rayleigh scattering for potential non-line-of-sight links. Next steps include testing longer distances, non-line-of-sight performance, and environmental effects on transmission.

UV / IR

PureFize Presents Field-emission Cathodoluminescent Approach for Next-generation Far-UVC Devices

Swedish UV technology company presents breakthrough results showing deep-UV emission below 240 nm from a scalable, phosphor-tunable device platform.

UV / IR

Violumas Has Launched New 255nm, 265nm, and 275nm LEDs

Violumas, provider of high-power UV LED solutions and inventor of 3-PAD LED technology, has released its next-generation 255nm, 265nm, and 275nm LEDs in both SMD and COB configurations.

UV / IR

LEDINTRO Introduces Breakthrough Harmless 230nm UV LED Air Sterilization Ball

Ultraviolet radiation has long been recognized as an effective method for inactivating airborne and surface-borne microorganisms. However, the widespread use of conventional UV-C sources has been constrained by safety concerns, as prolonged human exposure to traditional germicidal wavelengths is…

LuminusEdison optoLTF TechnologyLoreos

The latest Press Releases

Explore All

Stay current with our monthly Update from the lighting industry

LEDprofessional Newsletter

You will receive a confirmation e-mail. Unsubscribe anytime.